Simple & Multifocal IOLs
| |
|
Measurement of intraocular lenses (IOL) can performed with the high resolution PhaseView's Digital Wavefront Camera (DWC®) and the GetWave® software. The capability of DWC® - based measurement system for quality control of IOL's has been successfully tested through measurements of Sphere and Cylinder characteristics, for silicon and acrylic lenses, aspheric and biconvex surfaces. Typically, measurements with DWC® and GetWave® yield results within 0.5% of the designed values, for lenses with powers in the range 10D - 30D. Dedicated GetWave® software's interface is used to display results and control alignment and aberrations dynamically (at 15 Hz measurement rate). It allows display of intensity data, calculated wavefront and corresponding Zernike coefficients. |
![]() ![]() |